Model Number 470183-19
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/01/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a da vinci-assisted cholecystectomy procedure, the permanent cautery hook (pch) instrument broke into pieces.One piece was retrieved and two plastic pieces were noted to be missing.The customer confirmed that intraoperative imaging was performed to look for the remaining two fragments.Hence, the procedure was completed with a delay of approximately 2 hours.
 
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) received the permanent cautery hook (pch) instrument to perform failure analysis.The instrument was analyzed and found to have a broken monopolar yaw pulley at the distal clevis.Broken pieces are missing as a result of breakage.The size of the missing piece is approximately 5.96mm x 6.05mm.The components surrounding the break did not exhibit damage that would have contributed to the broken yaw pulley.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EndoWrist
Common Device Name
PERMANENT CAUTERY HOOK
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
+1(408)523-2100
MDR Report Key50873400
Report Number2955842-2026-39315
Device Sequence Number18809946
Product Code NAY
UDI-Device Identifier00886874112311
UDI-Public(01)00886874112311(10)K12260226
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K131861
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Operator of Device Health Professional
Device Model Number470183-19
Device Catalogue Number470183
Device Lot NumberK12260226 0002
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/16/2026
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 08/31/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/26/2026
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
da Vinci Instruments and Accessories
Outcome Attributed to Adverse Event Required Intervention;
Patient Age39 YR
Patient SexFemale
Patient Weight350 LB
-
-