| Model Number |
470183-19 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted cholecystectomy procedure, the permanent cautery hook (pch) instrument broke into pieces.One piece was retrieved and two plastic pieces were noted to be missing.The customer confirmed that intraoperative imaging was performed to look for the remaining two fragments.Hence, the procedure was completed with a delay of approximately 2 hours.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the permanent cautery hook (pch) instrument to perform failure analysis.The instrument was analyzed and found to have a broken monopolar yaw pulley at the distal clevis.Broken pieces are missing as a result of breakage.The size of the missing piece is approximately 5.96mm x 6.05mm.The components surrounding the break did not exhibit damage that would have contributed to the broken yaw pulley.
|
| |
|