MAUDE Adverse Event Report: Medtronic Puerto Rico Villalba OmniaSecure¿ MRI SureScan¿; PERMANENT DEFIBRILLATOR ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medtronic Puerto Rico Villalba OmniaSecure¿ MRI SureScan¿; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number 3930M79
Medical Device Problem Codes Failure to Capture (1081); Over-Sensing (1438); Decreased Sensitivity (2534); Capturing Problem (2891); Device Dislodged or Dislocated (2923); High Capture Threshold (3266)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the left bundle branch area pacing (lbbap) lead exhibited rising thresholds with no capture, p-wave oversensing (pwos) and decreased r waves.It was noted that the lbbap lead had dislodged.The lead was explanted and replaced.
 
Additional Manufacturer Narrative
Continuation of d10:  3501 lead, implanted (b)(6) 2023.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OmniaSecure¿ MRI SureScan¿
Common Device Name
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
Medtronic Puerto Rico Villalba
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
Medtronic Puerto Rico Villalba
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50875936
Report Number2649622-2026-28234
Device Sequence Number9224254
Product Code NVY
UDI-Device Identifier00763000643454
UDI-Public(01)00763000643454(17)270701(20)01
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P240036
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number3930M79
Device Catalogue Number3930M79
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2026
Initial Report FDA Received Date09/23/2026
Date Device Manufactured12/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DDPA2D4 ICD, 5076-52 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age69 YR
Patient SexFemale
-
-