| Model Number |
3501 |
| Medical Device Problem Code |
Over-Sensing (1438)
|
| Health Effect - Clinical Code |
Asystole (4442)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) and electrode stored an untreated episode that appeared show a true pause.Review of the episode shows signal double-counting was observed at times, with possible agonal rhythm or possible cardiopulmonary resuscitation (cpr) compressions before the pause event.The patient was currently intubated at the hospital.This electrode remains in service.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The local area field representative was contacted for additional information regarding event cause and resolution.At this time, no further information is available.Should additional information become available this report will be updated.This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|