• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EMBLEM? S-ICD; Implantable Cardioverter Defibrillator (Non-Crt)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EMBLEM? S-ICD; Implantable Cardioverter Defibrillator (Non-Crt) Back to Search Results
Model Number 3501
Medical Device Problem Code Over-Sensing (1438)
Health Effect - Clinical Code Asystole (4442)
Date of Event 08/26/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) and electrode stored an untreated episode that appeared show a true pause.Review of the episode shows signal double-counting was observed at times, with possible agonal rhythm or possible cardiopulmonary resuscitation (cpr) compressions before the pause event.The patient was currently intubated at the hospital.This electrode remains in service.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
The local area field representative was contacted for additional information regarding event cause and resolution.At this time, no further information is available.Should additional information become available this report will be updated.This product investigation is still in progress.This report will be updated when product investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM? S-ICD
Common Device Name
Implantable Cardioverter Defibrillator (Non-Crt)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
4100 hamline avenue north
saint paul MN 55112 -579
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50876649
Report Number2124215-2026-50837
Device Sequence Number10716284
Product Code LWS
UDI-Device Identifier00802526599200
UDI-Public(01)00802526599200(17)200321
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P110042
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/21/2020
Device Model Number3501
Device Catalogue Number3501
Device Lot Number117303
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2026
Initial Report FDA Received Date09/24/2026
Date Device Manufactured03/21/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age61 YR
Patient SexFemale
-
-