| Model Number |
HexaVue IP |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported during a patient procedure the monitor to their hexavue ip custom room rack was experiencing a routing issue that attributed to the surgical display images ¿dropping¿ and ¿going black¿.No report of procedure delay.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite to inspect the hexavue ip custom room rack and found the bs2 switch needed to be rebooted.To resolve the issue the technician rebooted the bs2 switch, tested the function and operation of the unit, confirmed it to be operating to specification, and returned the unit to service.No additional issues have been reported.
|
| |
|