• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR INC. Perclose¿ ProStyle¿; Device, Hemostasis, Vascular

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR INC. Perclose¿ ProStyle¿; Device, Hemostasis, Vascular Back to Search Results
Model Number 12773-03
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported this was an arteriotomy closure of the calcified right common femoral artery using the pre-close technique prior to a transcatheter aortic valve replacement (tavr) procedure.The sutures of two prostyle devices were successfully pre-placed.The sheath was upsized to greater than 10f and the tavr procedure was completed.Reportedly post procedure, one of the blue rail sutures broke when tightening down the suture over the wire using the snare knot pusher.Enough of the rail suture that broke was left to advance the knot successfully and achieve hemostasis with that same suture and the other successfully pre-placed suture.There were no reported adverse patient effects.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the reported information, the reported suture separation could not be determined.Factors that may contribute to suture separation during knot advancement include, but are not limited to, user technique (tensioning angle not coaxial) or suture/ nick damage.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Perclose¿ ProStyle¿
Common Device Name
Device, Hemostasis, Vascular
Manufacturer (Section D)
ABBOTT VASCULAR INC.
3200 lakeside drive
santa clara CA 95054 -280
Manufacturer (Section G)
ABBOTT VASCULAR, Reg # 3005718570 (P099)
cashel road
clonmel tipperary UNK
EI   UNK
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591--462
+1(951)914-5259
MDR Report Key50876815
Report Number2024168-2026-03684
Device Sequence Number9223796
Product Code MGB
UDI-Device Identifier08717648235184
UDI-Public(01)08717648235184(17)280430(10)6050643
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P960043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number12773-03
Device Catalogue Number12773-03
Device Lot Number6050643
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/06/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Closure Device: ProStyle
Patient SexUnknown
-
-