• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Olympus Winter & Ibe GmbH Berlin Facility HF unit "ESG-400"; Electrosurgical generator ESG-400

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Olympus Winter & Ibe GmbH Berlin Facility HF unit "ESG-400"; Electrosurgical generator ESG-400 Back to Search Results
Model Number WB91051W
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the electrosurgical generator esg-400 displayed an e433 error, and the system is restarting automatically.The issue occurred during set up or inspection for use.There were no reports of patient harm.
 
Additional Manufacturer Narrative
E1 - initial reporter establishment name: (b)(6).E1 - address: (b)(6).The device was returned to olympus for inspection.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the most likely cause of failure is an electrical defect, such as electronic components damage or deterioration, poor contact at electrical connection points, etc.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF unit "ESG-400"
Common Device Name
Electrosurgical generator ESG-400
Manufacturer (Section D)
Olympus Winter & Ibe GmbH Berlin Facility
rheinstrasse 8
teltow, brandenburg 14513
GM  14513
Manufacturer (Section G)
Olympus Winter & Ibe GmbH Berlin Facility
rheinstrasse 8
teltow, brandenburg 14513
GM   14513
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50876826
Report Number3003724334-2026-00502
Device Sequence Number15759113
Product Code GEI
UDI-Device Identifier04042761076838
UDI-Public04042761076838
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K203682
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWB91051W
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/08/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/27/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/08/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1