• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Fisher & Paykel Healthcare Ltd Fisher & Paykel Healthcare; Respiratory Humidifier

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Fisher & Paykel Healthcare Ltd Fisher & Paykel Healthcare; Respiratory Humidifier Back to Search Results
Model Number PT101US
Medical Device Problem Code Defective Alarm (1014)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A healthcare facility in canada reported that the power out alarm of the pt101 airvo 2 humidifier (airvo 2) sounded for less than 120 seconds.This was found while the subject airvo 2 was out of use and not on a patient.This was reported subsequent to a field safety notice informing users of a change to the disinfection kit user manual of the airvo 2 to include a regular test of the power out alarm, to be carried out in between each patient use.There was no patient involvement as the issue was found whilst the device was not in use on a patient.
 
Additional Manufacturer Narrative
(b)(4) product background: the pt101 airvo 2 humidifier (airvo 2) is an active humidifier with integrated flow generator that delivers high flow warmed and humidified respiratory gases to spontaneously breathing patients through a variety of patient interfaces.Its intended use is for the treatment of spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases.The airvo 2 is not intended for life support, and appropriate patient monitoring must be used at all times.Method: the subject airvo 2 was not returned to fisher & paykel healthcare (f&p) for evaluation.F&p's investigation is based on the information provided by the healthcare facility, previous investigations of similar complaints, and f&p's knowledge of the product.Results: the healthcare facility has stated that the power out alarm of the subject airvo 2 sounded for less than 120 seconds.There was no patient involvement as the issue was found whilst the device was not in use on a patient.Conclusion: based on f&p's investigation, knowledge of the product, and previous investigations of similar complaints, the power out alarm of the subject airvo 2 sounding for less than 120 seconds is likely due to the aging of component supercapacitors.A review of all-time complaint data has confirmed that globally there have been no events with serious injury or death related to the power out alarm sounding for less than 120 seconds.F&p's investigation found it likely that during the covid-19 pandemic, airvo 2 devices were kept continuously connected to mains power to ensure immediate availability, given the urgency and unpredictability of respiratory deterioration in covid-19 patients.This practice aligns with hospital protocols, which emphasize readiness and rapid deployment of respiratory support equipment in high-acuity settings.The original life testing carried out on airvo 2 devices assumed that the devices were disconnected from mains between therapy sessions and as practices have evolved, this no longer reflects real world usage patterns, particularly during pandemic conditions when device availability was critical and since clinical practices have shifted.F&p's investigation also assessed the manufacturing controls and determined they were sufficient and adequate.In addition, as the root cause pointed to usage in field, no further changes to the manufacturing or supplier processes was deemed necessary.During the manufacturing process, quality control measures ensure each manufactured airvo 2 meets specification.These quality control measures are performed at the end of the final assembly process on 100% of the manufactured units.Any unit that fails any of these tests is rejected.F&p have updated test protocols to specify that test devices be powered 24 hours a day i.E.Worst-case, and f&p expect this change to cover any further unexpected operational deviations or unanticipated changes in usage patterns.This update to f&p's test protocols will be used in any new development or change to the device.Furthermore, an update to the disinfection kit user manual of the airvo 2 was made to include a regular test of the power out alarm, to be carried out in between each patient use.This change was communicated to users through a voluntary field safety notice in september 2025.In response to this change, customers are conducting the regular test in between patient use and reporting any failures.The user instruction (ui) of the airvo 2 humidifier states: - "if either alarm signal is absent, do not use the unit.Contact your fisher & paykel healthcare representative." - "the unit is not intended for life support." - "appropriate patient monitoring must be used at all times.Loss of power means loss of therapy." - "use continuous oxygen monitoring on patients who would desaturate significantly in the event of disruption to their oxygen supply.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Fisher & Paykel Healthcare
Common Device Name
Respiratory Humidifier
Manufacturer (Section D)
Fisher & Paykel Healthcare Ltd
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
Fisher & Paykel Healthcare Ltd
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
+1(949)434-1005
MDR Report Key50876855
Report Number9611451-2026-02783
Device Sequence Number16131190
Product Code BTT
UDI-Device Identifier09420012422347
UDI-Public(01)09420012422347(10)2101075246(11)200416
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K131895
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberPT101US
Device Catalogue NumberPT101US
Device Lot Number2101075246
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/16/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
9611451-09/10/2025-001-C
Patient Sequence Number1
Patient SexUnknown
-
-