• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Gyrus ACMI, Inc. 200 MICRON TFL BALL TIP SINGLE USE FIBER; Powered Laser Surgical Instrument

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Gyrus ACMI, Inc. 200 MICRON TFL BALL TIP SINGLE USE FIBER; Powered Laser Surgical Instrument Back to Search Results
Model Number TFL-FBX200BS
Medical Device Problem Codes Sparking (2595); Flare or Flash (2942)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/28/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when the laser of the subject device fired, it sparked and appeared to ignite/burn.There were no reports of patient harm/involvement.During the initial evaluation of the device, the reported event was unable to be confirmed.
 
Additional Manufacturer Narrative
The device was inspected on site via field service, and the reported failure was not confirmed.Based on the results of the investigation, a root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
200 MICRON TFL BALL TIP SINGLE USE FIBER
Common Device Name
Powered Laser Surgical Instrument
Manufacturer (Section D)
Gyrus ACMI, Inc.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
Gyrus ACMI, Inc.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50876974
Report Number3011050570-2026-00933
Device Sequence Number15797709
Product Code GEX
UDI-Device Identifier00821925043985
UDI-Public00821925043985
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K242191
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTFL-FBX200BS
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/28/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-