• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical (CRM-Sylmar) DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; Permanent Defibrillator Electrodes

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical (CRM-Sylmar) DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; Permanent Defibrillator Electrodes Back to Search Results
Model Number 7122Q/65
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fever (1858); Unspecified Infection (1930)
Date of Event 08/28/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient was implanted with implantable cardioverter defibrillator (icd) on (b)(6) 2026.Post-operatively, the patient experienced recurrent fever.Subsequent echocardiography and ultrasound revealed vegetation on the atrial lead.A causal relationship between the reported infection and the procedure or the abbott device could not be determined.Despite prolonged antibiotic therapy, the patient's condition did not improve.As a result, the icd, atrial lead and right ventricle (rv) lead were explanted on (b)(6) 2026.Blood culture was taken however, the result was not reported.The patient had no consequences throughout the procedure.The cause of infection was related to vegetation on atrial lead.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Common Device Name
Permanent Defibrillator Electrodes
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50877066
Report Number2017865-2026-20254
Device Sequence Number9223806
Product Code NVY
UDI-Device Identifier05414734503341
UDI-Public(01)05414734503341(10)A000186138(17)290331
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
P950022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number7122Q/65
Device Lot NumberA000186138
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/07/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-