The customer explained that the cuffs are not holding air.We received 1 ea (b)(6) handheld pvc aneroids; cuff was inspected for any holes, and no damage was observed to cuff, bulb, tubing, or dial.Testing was then performed by inflating cuff to 300mmhg; after cuff was inflated, valve was opened to deflate, and leaking did not occur prior to opening valve.The complaint was not confirmed, and division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|