| Model Number |
RT266 |
| Medical Device Problem Codes |
Crack (1135); Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A distributor reported on behalf of a healthcare facility in japan, via a fisher & paykel (f&p) healthcare field representative, that an rt266 infant dual-heated evaqua2 breathing circuit failed the pre-use ventilator leak test due to a crack on the dome of the mr290v vented autofeed humidification chamber provided with the breathing circuit kit.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4)
the subject mr290v vented autofeed humidification chamber provided with the rt266 infant dual-heated evaqua2 breathing circuit has been returned to fisher & paykel (f&p) healthcare in new zealand for evaluation.A follow up report will be provided upon completion of our investigation.
|
| |
|