Effective august 1, 2026, mdr-related activities and responsibilities were transferred to pentax medical corporation as part of a corporate restructuring.This report is being submitted by pentax medical corporation.In accordance with 21 cfr 803.3(m), because this corporate restructuring does not affect the manufacturing site, the fda registration number previously assigned for this site (#9610877) continues to be used.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the distal end with ccd module/us probe blackout.Based on the result, we concluded that it was caused due to the excessive force applied on the distal end with ccd module/us probe.In addition, our technician confirmed that the insertion flexible tube fluid damage, the segment fluid damage, the control body fluid damage, the light guide cable fluid damage, the ccd drive pcb fluid damage, the us connector cable (long) fluid damage, the ultrasound connector body fluid damage, the lg connector fluid damage, the us connector cable (long) buckled, the control body corroded, the ccd drive pcb corroded, the ultrasound connector body corroded, the lg connector corroded, the shield cover corroded, the operation channel (primary) leak, the operation channel (primary) cut, the remote control button(1) cut, and the us probe blackout; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
|