• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Hamilton Medical AG HAMILTON-C3

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Hamilton Medical AG HAMILTON-C3 Back to Search Results
Model Number 160005
Medical Device Problem Code Loss of Power (1475)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/12/2026
Type of Reportable Event Malfunction
Event or Problem Description
Hamilton medical ag received the following event description: biomed states that the device shows loss of external power.Says it happens intermittingly.Tried different power cord and different outlet.Issue remains.Biomed opened device to look for possible loose connections, none found.- no health consequences or impact - no intervention necessary.
 
Additional Manufacturer Narrative
Hamilton medical ag ref.Nr: (b)(4).Investigation is still ongoing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAMILTON-C3
Common Device Name
HAMILTON-C3
Manufacturer (Section D)
Hamilton Medical AG
via crusch 8
bonaduz 7402
SZ  7402
Manufacturer Contact
ann franz
via crusch 8
bonaduz 7402
SZ   7402
MDR Report Key50877479
Report Number3001421318-2026-00593
Device Sequence Number9379983
Product Code CBK
UDI-Device Identifier07630002801881
UDI-Public(01)07630002801881(11)201012(240)160005
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number160005
Device Catalogue Number160005
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2026
Initial Report FDA Received Date09/24/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-