• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY? XD; coronary drug-eluting stent

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY? XD; coronary drug-eluting stent Back to Search Results
Model Number H7493941820220
Medical Device Problem Codes Device-Device Incompatibility (2919); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that stent dislodgement occurred.The patient presented with coronary artery disease and underwent percutaneous coronary intervention.The 80% stenosed target lesion was located in the moderately tortuous and moderately calcified right coronary artery.During the procedure, a 7f guide catheter and a guidezilla ii guide extension catheter were used.After the lesion was pre-dilated with a 2.00mm emerge balloon catheter, a 2.25 x 20mm synergy? xd was advanced for treatment.During insertion, the stent got caught on the guidezilla and started to dislodge from the delivery system.The physician removed the delivery system, and guide extension together, retrieving the dislodged stent with no stent left behind.The guidezilla was reinserted, and a new stent was then introduced and successfully deployed.No patient complications were reported.
 
Additional Manufacturer Narrative
Good faith effort attempts were made to try and retrieve additional details regarding the reported event, but further information was unable to be obtained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY? XD
Common Device Name
coronary drug-eluting stent
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
ballybrit business park
galway H91 Y 868
EI   H91 Y868
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
+1(612)875-7234
MDR Report Key50877502
Report Number2124215-2026-50842
Device Sequence Number9380502
Product Code NIQ
UDI-Device Identifier08714729980735
Combination Product (Y/N)Y
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH7493941820220
Device Catalogue NumberH7493941820220
Device Lot Number0039065088
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2026
Initial Report FDA Received Date09/24/2026
Date Device Manufactured03/31/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-