It was reported that stent dislodgement occurred.The patient presented with coronary artery disease and underwent percutaneous coronary intervention.The 80% stenosed target lesion was located in the moderately tortuous and moderately calcified right coronary artery.During the procedure, a 7f guide catheter and a guidezilla ii guide extension catheter were used.After the lesion was pre-dilated with a 2.00mm emerge balloon catheter, a 2.25 x 20mm synergy? xd was advanced for treatment.During insertion, the stent got caught on the guidezilla and started to dislodge from the delivery system.The physician removed the delivery system, and guide extension together, retrieving the dislodged stent with no stent left behind.The guidezilla was reinserted, and a new stent was then introduced and successfully deployed.No patient complications were reported.
|