| Model Number |
380721-09 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted retroperitoneal anterior partial nephrectomy surgical procedure, an intuitive surgical (is) technical support engineer (tse) was contacted for by the customer to report that the system shuts down by itself, mid procedure.As system was not onsite connected, no logs were available, tse asked customer to take pictures on the event logs in the touchscreen of the vision side cart (vsc), but nothing received yet.No idea what triggered the shut down or on which surgeon side console (ssc).A field service engineer (fse) would need to go on site to upload the logs so they can be analyzed to see what are the recommended actions.System was hard power cycled due to other problems.The procedure was completed with no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was completed to determine the cause of this reported event.An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.Fse replaced the auxiliary video board (avp) to correct the reported event.The system was tested and verified as ready for use.Isi did not receive a da vinci product involved with this complaint to perform failure analysis.The complaint was confirmed based on the field evaluation, which indicates that the device may have contributed to the customer reported issue.
|
| |
|