Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The endoscope plus instrument was analyzed and found with damage at the distal end.The distal window located at distal tip was visually inspected and found cracked.Additionally, the endoscope was visually inspected and damage was found at the distal end.The distal window located at the distal tip was visually inspected and found to have a broken or detached piece in a location that could fall into the patient.The endoscope was tested and placed on in-house system for functional testing and failed to provide a video due to an electrical communication failure.The in-house system or equivalent failed to communicate with the endoscope, resulting in an error message: check endoscope cable connection/endoscope self - test failed.The endoscope was visually inspected and found with mechanical damage the scope¿s distal tip.The endoscope was tested and placed on in-house system for functional testing and failed to endoscope diagnostic test (edt) not recognize.The endoscope was evaluated and found with a camera instrument adapter defect.The endoscope was placed through friction test using a screening aide to manually rotate the adapter to detect for friction.The camera instrument adapter did not rotate properly and/or noises were heard during testing and confirmed as binding.The camera instrument adapter removed from endoscope housing and evaluated and found with attached endoscope adapter (aea) shaft bearing contributing to friction.The endoscope was tested and place on an in-house system for cable testing and failed due to wpb defect.The probable root cause of the cracked lens is attributed to use conditions or improper handling.Accidental drops of the endoscope, inadvertent collisions with hard surfaces, or improper cleaning during reprocessing can result in damaged optical components.A cracked window can also result in fluid invasion that may further the damage to optical components.
|