It was reported that during a da vinci assisted ureteroplasty surgical procedure, the sheath indicator on the energy shield monitor (esm) displayed yellow with a warning related to the instrument sheath/cord.The procedure was completed as planned with no patient injury.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: the customer noted the procedure was continued and completed despite the yellow cord/sheath indicator without using monopolar, with bipolar and scissors.The monopolar device kept breaking down, so it was not available throughout the procedure.Colleagues had contacted phone support and tried everything they could think of; three cables were replaced, a new device was installed, and the system was completely shut down, etc.No alternative energy source, different instrument, or different system configuration of the sort was used to complete the procedure.
|
|
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.The fse went on-site and replaced the energy shield monitor (esm).The system was verified and ready for use.Intuitive surgical, inc.(isi) has not received the da vinci product with an alleged issue to perform failure analysis.A follow-up mdr will be submitted if the product is returned and evaluated and/or if additional information is received.Blank mdr fields:
the missing patient information in sections a and b was either unknown, unavailable, not provided, or not applicable.The expiration date for section d4 is not applicable.Field d6 is blank because the product is not implantable.Information for the blank fields in section e1 is not available.Fields g5 and g7 are not applicable.Field h10 is blank as there are no related report numbers.
|