| Model Number |
901001 |
| Medical Device Problem Codes |
Thermal Decomposition of Device (1071); Sparking (2595)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Baxter received a report stating that a power management stand emitted sparks and burning smell.The power cable appeared to be blackened.The process step during which this occurred was unknown.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The baxter technician found the power cord was defective.The reported problem has been verified.The baxter technician arranged to get the power cord replaced for the customer to resolve the issue.Although there was no injury associated with this event, if the reported event were to recur it could lead to injury or death.Therefore, baxter is reporting this complaint.
|
| |
|