| Model Number |
IntelliVue Patient Monitor MX450 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the monitor's non-invasive blood pressure (nbp) measurement frequency setting kept changing on its own from "auto" to "sequence" mode, forcing clinical staff to manually correct the interval with every single patient interaction.The device was in use on a patient at the time of event, there was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis was performed by a philips remote service engineer (rse) which included interviewing and assisting the customer with evaluating the unit.The rse included the philips clinical support engineer (cse) on the call and walked the customer through making a configuration change related to the nbp settings.The operation was then verified following the configuration change.The configuration was updated with assistance from the cse.Results of the monitor configuration are now set to nbp auto q5 min.The monitor is now performing as expected and the issue is resolved.Based on the information available and the case, the cause of the reported problem was incorrect configuration.The reported problem was confirmed.If additional information is received, the complaint file will be reopened for further investigation.
|
| |
|