• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Cochlear Ltd Nucleus CI612 Cochlear Implant with Contour Advance Electrode; Nucleus 24 Cochlear Implant System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Cochlear Ltd Nucleus CI612 Cochlear Implant with Contour Advance Electrode; Nucleus 24 Cochlear Implant System Back to Search Results
Model Number CI612
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/23/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Per the clinic, the device was explanted on (b)(6) 2026 due to an unspecified reason.Additional information has been requested but has not been made available as of the date of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Nucleus CI612 Cochlear Implant with Contour Advance Electrode
Common Device Name
Nucleus 24 Cochlear Implant System
Manufacturer (Section D)
Cochlear Ltd
1 university avenue
macquarie university, nsw 2109
AS  2109
Manufacturer (Section G)
Cochlear Ltd
1 university avenue
macquarie university, nsw 2109
AS   2109
Manufacturer Contact
amirah rosly
unit ug-1, vertical podium
no. 8 jalan kerinchi
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key50878067
Report Number6000034-2026-03354
Device Sequence Number13089994
Product Code MCM
UDI-Device Identifier09321502036818
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P970051
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date10/16/2025
Device Model NumberCI612
Device Catalogue NumberP774600
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/20/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/31/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-