| Model Number |
2C8519 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During evaluation of a clearlink system continu-flo solution set, leakage was observed due to detachment of the tubing from the clearlink during pressure testing.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.Visual inspection of the returned sample did not identify any abnormalities that could have contributed to the reported condition.Pressure testing under water was performed, during which a leak was observed due to detachment at the junction between the tubing and the clearlink luer activated y site of the second subassembly.The sample detached during the underwater leak test.Clear passage testing was also performed and yielded satisfactory results.The reported condition was verified.The cause of the issue was related to manufacturing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|