| Model Number |
500001002502 |
| Medical Device Problem Code |
Installation-Related Problem (2965)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medstation¿ es auxiliary, tower required maintenance when recover storage space.The customer stated that there was a delay in patient care. however, there were no adverse events or injuries reported based on this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 16-mar-2023 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that auxiliary tower was the source of the reported drawer failure condition.The field service engineer (fse) connected the auxiliary tower via the pyxisbus cable and assisted pharmacy staff with unloading medications from tower doors 1 and 2.Shut down the medstation, disconnected the old tower, and allowed the customer to delete the tower from medapplication.After removal of the old tower configuration, shut down the medstation again and connected the replacement 4-door auxiliary tower.The system functioned as intended after the field service engineer (fse) resolved the issue.
|
| |
|