It was reported that when using the bd pyxis¿ es refrigerator helmer refrigerator bins 1.1, 1.2, and 1.3 failed and displayed a device not on bus error.The customer rebooted the station and performed a recover storage space procedure; however, the issue persisted and the bins remained non-functional.The customer stated that this malfunction occurred while dispensing the medication which caused a delay to the patient care.There were no adverse events or injuries reported based on this event.
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 21-may-2025 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the refrigerator bins had failed and were not detected on the communication bus.The field service engineer (fse) inspected the unit and found refrigerator bins 1.1, 1.2, and 1.3 in a failed state.The field service engineer (fse) powered down the refrigerator, removed the top panel, replaced the refrigerator controller assembly, and configured the required jumper settings.After restarting the unit, all bins recovered successfully within the application.The customer tested the refrigerator bins and verified proper functionality.The system functioned as intended after field service engineer repaired the device.
|