| Model Number |
6640440 |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the ventilator failed internal leakage test with a message 'system volume too large' during pre-use check.There was no patient involvement.Manufacturer's ref.#: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A third party field service engineer (fse) has investigated the reported ventilator on-site.The air gas module was found defective.The device's logs were not provided for further analysis.The air gas module regulates the inspiratory gas flow and a faulty air gas module may affect the delivered volume or gas mixture (o2/air).The cause of the reported issue in this customer product complaint has been determined.The issue was related to air gas module.No further investigation can be done as the defective air gas module will not be returned for investigation.
|
| |
|