This event was also reported to fda by the importer under importer report no.2438477-2026-00058.(1) investigation and analysis methodology.The mechanism of electrotherapy devices may cause skin burns in patients, including joule heating burns and chemical burns.Potential causes include device-related issues (e.G., substandard electrodes, insufficient electrode-to-skin contact leading to concentrated current density, device malfunction causing excessive output current, excessive dc component per unit area combined with prolonged continuous treatment), operation-related issues (e.G., excessive treatment intensity, prolonged continuous treatment, adjusting electrodes before the device is stopped, electrodes placed too close together), and patient-related factors such as potential allergy to the gel.According to the customer's feedback, the patient has a history of long-term medication use and a documented history of neuromuscular disorder, with no neuropathy.Among the long-term medications, pregabalin and clonidine have analgesic effects, and pregabalin is explicitly associated with adverse symptoms including hypoesthesia; allegra (fexofenadine), as an antihistamine, is used for allergic rhinitis, urticaria, etc.The medications taken made the patient less sensitive to pain, which may have led to an excessively high treatment intensity setting, and heat accumulation during prolonged treatment may have caused the burn.Although the electrodes were new and used for the first time, a causal association with incomplete electrode-to-treatment-site contact and uneven current distribution of the electrodes themselves cannot be excluded.From the patient's burn images, the burn area (including redness and a central ulcerated area) is concentrated in the central region of the electrode, which is consistent with the causal hypothesis of localized current concentration forming a hot spot.(2) failure mode(s)/mechanism(s) and associated component(s).Based on the analysis of potential burn mechanisms, the burn cause has a strong causal association with substandard electrodes (uneven current density), insufficient electrode-to-skin contact during treatment (excessive current density per unit area), and the patient's reduced pain sensitivity.(3) conclusions: through the burn mechanism analysis, this burn has a strong causal association with substandard electrodes, insufficient electrode-to-skin contact during treatment, and the patient's reduced pain sensitivity.However, the possibility of device malfunction cannot be entirely excluded.As the customer has indicated that the device will be returned, it is recommended that the distributor perform output testing using the parameters used during the patient's treatment to fully exclude the possibility of device malfunction.The relevant information has been incorporated into the complaint file for traceability and fda review.
|