C.R. Bard, Inc. (BASD) -3006260740 PowerPort ClearVUE Slim Implantable Port, ChronoFlex Single-Lumen, Kit, 6F; Port & Catheter, Implanted, Subcutaneous, Intravascular
|
Back to Search Results |
|
| Catalog Number |
1616070 |
| Medical Device Problem Codes |
Reflux within Device (1522); Improper or Incorrect Procedure or Method (2017); Malposition of Device (2616)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent a port placement procedure using the powerport clearvue slim.During the procedure, an arterial puncture occurred.The procedure was discontinued after the arterial puncture was identified while a micropuncture kit and sheath introducer from another manufacturer were being used.The arterial puncture was managed with pressure hemostasis and did not require any additional medical or surgical intervention.It was further reported that a significant backflow of blood was observed when the catheter was inserted into the other manufacturer's sheath, leading to the recognition of the arterial puncture.It was also reported that, prior to the puncture, a pocket had been created for port placement, resulting in a scar at the pocket site.A picc was placed at a later date.The current status of the patient is healthy.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records will be performed.The sample has not been returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) are adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow up report will be submitted as applicable.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|
|
|