It was reported by the customer that during clinical use the cs300 intra-aortic balloon pump (iabp) experienced a blood back.The device shutdown due to blood back during therapy.There was patient involved, no harm reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported that after performing troubleshooting it was verified that the blood entered into unit through a hole in the catheter.The fse replaced all contaminated parts: tubing, fill manifold assembly, clamp cable, reservoir assembly, condensate removal module, pneumatic connector, purge valve assembly and safety disk.The fse conducted functional and safety testing.Product passed all functional and safety tests per manufacturer specifications.
|