Product event summary: the delivery system was returned without the device intact in the device cup, and analyzed.Analysis indicated the lumen of the delivery system was torn.The delivery system tether was frayed.The delivery system tether was knotted.The analyst noted the deployment test cannot be performed due to the device was not returned, and the tether and pin were not intact in the delivery system.Visual inspection showed the lumen torn, the tether was knotted and frayed.These finding positively contributed to the deployment of the device during implant.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|