Medtronic Ireland Micra¿ AV2; LEADLESS PACEMAKER Back to Search Results
Model Number MC2AVR1-DELSYS
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/10/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the implant procedure of a leadless implantable pulse generator (ipg) the delivery system exhibited a de ployment issue.The deployment button was broken and the leadless ipg was unable to be deployed.It was noted that when the physician attempted to deploy the device the deployment button moved but the device remained completely seated inside the cup.The delivery system was removed and it was noted that the leadless ipg was stuck inside the catheter at the valve.The leadless ipg was removed from the catheter by pulling on the tether.The leadless ipg was attempted/ not used and replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the delivery system was returned without the device intact in the device cup, and analyzed.Analysis indicated the lumen of the delivery system was torn.The delivery system tether was frayed.The delivery system tether was knotted.The analyst noted the deployment test cannot be performed due to the device was not returned, and the tether and pin were not intact in the delivery system.Visual inspection showed the lumen torn, the tether was knotted and frayed.These finding positively contributed to the deployment of the device during implant.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Micra¿ AV2
Common Device Name
LEADLESS PACEMAKER
Manufacturer (Section D)
Medtronic Ireland
parkmore business park west
galway
EI 
Manufacturer (Section G)
Medtronic Ireland
parkmore business park west
galway
EI  
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50880679
Report Number9612164-2026-04800
Device Sequence Number18582493
Product Code PNJ
UDI-Device Identifier00763000078645
UDI-Public(01)00763000078645(17)270814
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P150033
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC2AVR1-DELSYS
Device Catalogue NumberMC2AVR1-DELSYS
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/16/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/23/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age58 YR
Patient SexFemale
-
-