| Model Number |
0998-00-0800-45 |
| Medical Device Problem Code |
Inability to Auto-Fill (1044)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the customer that during use, the cardiosave intra-aortic balloon pump (iabp) malfunctioned.Autofill failure has been reported.No harm reported.When the machine malfunctioned, a new machine was installed, and the patients were not affected.The problem was found to be a malfunction in the inflation valve assembly.Field service engineer replaced the drive valve assembly; the machine system passed inspection and inflation was normal.The equipment passed all the inspections required by the factory.The equipment has been delivered to the customer and is ready for clinical use.
|
| |
|
Additional Manufacturer Narrative
|
|
Event site name: (b)(6)
event site postal code: 200120.Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|