| Catalog Number |
CV-4-15 |
| Medical Device Problem Codes |
Difficult to Advance (2920); Physical Resistance/Sticking (4012); Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during treatment for ¿hypo take down,¿ a concerto versa coil was prepared as indicated in the instructions for use and tracking difficulty occurred during initial advancement and resistance/sticking occurred in the catheter at the hub. physician hubbed introducer sheath to catheter and felt resistance advancing into a 2.8 non-medtronic microcatheter. physician pulled back and tried to re-advance and felt resistance again.When physician pulled coil back , the coil had prematurely detached inside the microcatheter. the pusher release wire was bent/broken. the detached coil was removed from the patient.A replacement concerto versa cv-3-15 coil was used to complete the procedure and the issue was resolved.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|