| Model Number |
ARDVST229039A |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a cesarean section, while adjusting the position of the surgical light, the rear cover detached from the spring arm of the volista standop surgical light due to significant mechanical contact and fell onto the sterile table during the operation, resulting in contamination of the sterile instruments.After re-preparing the sterile instruments, the surgery continued, and the procedure time was extended.The issue was confirmed based on photographic evidence.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Event site name: (b)(6).An investigation was conducted regarding the detachment of the rear cover from the spring arm and its fall onto the sterile table.Evaluation of the provided evidence showed detachment consistent with significant mechanical contact during routine operating room handling.No evidence was identified to suggest a manufacturing, material, or design-related defect.The associated risk of particulate generation and potential contamination of the sterile field has been previously identified and assessed within the risk management file, and existing mitigations remain effective.This issue is managed within the manufacturer¿s quality system and monitored through ongoing trending activities.No capa or field action is required at this time.
|
| |
|