Model Number 160021
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2026
Type of Reportable Event Malfunction
Event or Problem Description
Hamilton medical ag received the following event description: upon starting operation of the hamilton-c6, the device fails to reach the programmed tidal volume.The pre-use check procedure was optimized, and the unit was restarted with new consumables, but the issue persisted.Active humidification settings were also adjusted¿considering that excessive condensation might be interfering with measurements¿but this did not resolve the problem either.Despite optimizing the setup of consumables, the pre-use check, and humidification settings, the issue remained unresolved.The replacement and verification of the complete patient circuit were reported as the medical intervention performed.No harm to the patient was reported.
 
Additional Manufacturer Narrative
Hamilton medical ag reference number: (b)(4) investigation ongoing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAMILTON-C6
Common Device Name
HAMILTON-C6
Manufacturer (Section D)
Hamilton Medical AG
via crusch 8
bonaduz 7402
SZ  7402
Manufacturer Contact
alina liliana valean
via crusch 8
bonaduz 7402
SZ   7402
MDR Report Key50881637
Report Number3001421318-2026-00596
Device Sequence Number18807544
Product Code CBK
UDI-Device Identifier07630002808590
UDI-Public(01)07630002808590(11)190405(240)160021
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
K201658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Distributor
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number160021
Device Catalogue Number160021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2026
Initial Report FDA Received Date09/24/2026
Date Device Manufactured04/05/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-