• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Laboratories Alinity i Processing Module; ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Laboratories Alinity i Processing Module; ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE Back to Search Results
Catalog Number 03R65-01
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer observed falsely elevated alinity i free t4 results for multiple patients.Results were released from the laboratory.Same sample was used for the repeat testing.The following results were provided: (customer provided reference range: : 9.00 to 20.00 pmol/l).(b)(6) 2026, sid (b)(6); initial result => 64.35 pmol/l, repeat results on (b)(6) 2026 (b)(6)= 19.72 pmol/l and 11sep2026= 20.22 pmol/l.(b)(6) 2026, sid: (b)(6); initial result = 22.55 pmol/l, repeat results on (b)(6) 2026 (b)(6)= 13.27 pmol/l.(b)(6) 2026, sid: (b)(6); initial result = 24.48 pmol/l, repeat results on (b)(6) 2026 (b)(6)= 13.18 pmol/l.No additional patient information was available.No impact to patient management was reported.
 
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Section a patient information: refer to section b5 for complete patient information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Alinity i Processing Module
Common Device Name
ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
Manufacturer (Section D)
Abbott Laboratories
1915 hurd drive
irving TX 75038
Manufacturer (Section G)
Abbott Laboratories
1915 hurd drive
irving TX 75038
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
+49()6122582960
MDR Report Key50881801
Report Number3016438761-2026-00532
Device Sequence Number10754493
Product Code JJE
UDI-Device Identifier00380740137366
UDI-Public(01)00380740137366
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K170317
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03R65-01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALNTY I FREE T4 1200T, 07P70-30, 88073UD00
Patient SexUnknown
-
-