The account reported finding open seal.A sample was received for investigation; finished good production records were reviewed and there were no run without approvals, deviations or non-conformance reports that might have had an impact on this issue, the finished good pack build was reviewed and there were no issues with the build that likely contributed to the reported issue, and trending found 0 additional reported complaint(s) for this issue for the account in the past 6 months.The reported problem was confirmed: the pack received was opened on the top and did not appear to be sealed at all, and no other defects or damage was observed.Based on the complaint details this issue was likely caused by a manual error in the sealing process during pack assembly; production and qc managers were notified, feedback will be provided to the build team for awareness and to prevent recurrence in the future, and the issue will be monitored for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|