| Catalog Number |
ESBF3614C103E |
| Medical Device Problem Codes |
Leak/Splash (1354); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
An endurant iis bifurcated stent graft system was implanted during the treatment of a 77mm abdominal aortic aneurysm.It was reported approximately 1 month post the index procedure, ct imaging identified a type ia endoleak and stent graft infolding midway through the main body.Per the physician the cause of the event was undetermined and the event is unresolved.No additional clinical sequelae were reported and the patient is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|