• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Black Forest Medical GmbH DORO® Skull Clamp; DORO® QR3 Skull Clamp

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Black Forest Medical GmbH DORO® Skull Clamp; DORO® QR3 Skull Clamp Back to Search Results
Model Number 1001.001
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code Laceration(s) (1946)
Date of Event 08/24/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Customer informed us, that one of our products was involved in a case in which the patient sustained a cut to the scalp.
 
Additional Manufacturer Narrative
Based on the findings of the investigation, the deviation identified (stiffness of the two-pin holder) presumably did not cause or contribute to the described incident.The skull clamp sent in was combined with a torque screw from a third-party manufacturer.It is therefore not possible to verify the conformity of the third-party product (clamping force verification).Risks due to impermissible combinations (third party torque screw) cannot be ruled out as a contributing factor to adverse events such as a loss of clamping force.In addition, the customer informed us, that the doctor may not have locked the open/lock-mechanism of the clamp properly.The customer is referred to the relevant instructions in the instructions for use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORO® Skull Clamp
Common Device Name
DORO® QR3 Skull Clamp
Manufacturer (Section D)
Black Forest Medical GmbH
boetzinger strasse 86
freiburg im breisgau, baden-wuerttemberg 79111
GM  79111
Manufacturer (Section G)
Black Forest Medical GmbH
boetzinger strasse 86
freiburg im breisgau, baden-wuerttemberg 79111
GM   79111
Manufacturer Contact
julia hundertmark
boetzinger strasse 86
freiburg im breisgau, baden-wuerttemberg 79111
GM   79111
MDR Report Key50881986
Report Number3003923584-2026-00025
Device Sequence Number9939280
Product Code HBL
UDI-Device Identifier04250435506196
UDI-Public04250435506196
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K203505
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Remedial Action Repair,Replace
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1001.001
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/03/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/17/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-