• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS MEXICO, S. de R.L. de C.V. AMSCO; Sterilizer

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS MEXICO, S. de R.L. de C.V. AMSCO; Sterilizer Back to Search Results
Model Number AMSCO 400 Steam Sterilizer
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/25/2026
Type of Reportable Event Malfunction
Event or Problem Description
The user facility reported that their amsco 400 sterilizer was leaking water and emitting a ¿burning odor¿.No report of injury.
 
Additional Manufacturer Narrative
A steris service technician arrived onsite following the reported event to inspect the amsco 400 sterilizer and found a generator control contactor stuck closed.As the contactor was stuck in the closed position, this allowed continuous power to be provided to the unit causing excess steam to build up.As designed, the safety valve opened, and excess steam emitted from the unit.To resolve the issue, the technician replaced the contactor and necessary components.After the unit was tested and confirmed operating according to specifications it was returned to service.A 2-year complaint review confirmed this to be an isolated event.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMSCO
Common Device Name
Sterilizer
Manufacturer (Section D)
STERIS MEXICO, S. de R.L. de C.V.
avenida avante 790
parque industrial guadalupe
guadalupe nuevo leon 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. de R.L. de C.V.
avenida avante 790
parque industrial guadalupe
guadalupe nuevo leon 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
+1(440)259-7401
MDR Report Key50882026
Report Number3005899764-2026-00074
Device Sequence Number9826071
Product Code FLE
UDI-Device Identifier00724995209421
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K212424
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAMSCO 400 Steam Sterilizer
Device Catalogue NumberSR0201110111
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/02/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-