There was an allegation of questionable cobas® hiv-1 (192t) results with 3 patient samples tested on the cobas 6800 analyzer.It was reported that the customer was performing a lot to lot comparison and repeated 3 samples that were initially tested with a previous lot, n09424, and then with lot p03576.Sample 1:
on (b)(6) 2026, the initial result was 9.21e+01 iu/ml using lot n09424.On (b)(6) 2026, the repeat result was 2.17e+01 iu/ml using lot p03576
on (b)(6) 2026, the repeat result was 2.74e+01 iu/ml using lot p03576.Sample 2:
on (b)(6) 2026, the initial result was 3.86e+01 iu/ml using lot n09424.On (b)(6) 2026, the repeat result was negative using lot p03576.Sample 3:
on (b)(6) 2026, the initial result was negative using lot n09424.On (b)(6) 2026, the repeat result was negative using lot p03576.On (b)(6) 2026, the repeat result was
|