| Model Number |
M00546780 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Block h6:\
imdrf code a150101 captures the reportable event of detachment of device or device component\
imdrf code f23 captures the reportable event of unexpected medical intervention\
\
investigation summary\
the spy discover catheter was not returned.Therefore, the reported events of \"unretrieved device fragments (udf)\" and \"unexpected medical intervention\" could not be confirmed due to the lack of evidence.\
it was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.\
a similar complaint review was completed.There were no emerging trends identified that warrant further action.\
based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6:\
imdrf code a150101 captures the reportable event of detachment of device or device component;\
imdrf code f23 captures the reportable event of unexpected medical intervention.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a spy discover catheter was utilized in a lcbde (laparoscopic common bile duct exploration) procedure on (b)(6) 2026 for treatment of stones in the common bile duct.\
\
during the procedure, a piece of the light blue covering at the distal end of the spyglass discover catheter had chipped off into the patient abdomen.The physician performed an an exploratory laparotomy for small bowel resection however was not able to locate the piece that chipped off.Patient is scheduled for a follow up visit.\
\
procedure was completed with the original device.\
\
no other patient complications were reported as a result of the event.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a spy discoverer catheter was utilized in a lcbde (laparoscopic common bile duct exploration) with spyglass discover on (b)(6) 2026.\
\
during the procedure, a piece of the light blue covering at the distal end of the spyglass discover catheter had chipped off into the patient abdomen.The physician performed an an exploratory laparotomy for small bowel resection however was not able to locate the piece that chipped off.Patient is scheduled for a follow up visit.\
\
procedure was completed with the original device.\
\
no other patient complications were reported as a result of the event.
|
| |
|