The investigation determined that the operator sustained a thumb laceration during separate events involving the removal of metal seals from various vitros calibrator bottles.The operator did not seek medical attention or treatment for any of the injuries.No long-term serious injury was sustained.The assignable cause of this event is user error.However, based on a medical consultation with an ortho medical safety officer (mso) on (b)(6) 2026: per the sds, the affected calibrators contain human-derived materials, and no known test method can provide complete assurance that infectious agents are absent.Although the materials used in manufacturing are tested for known blood-borne pathogens, the possibility of untested, emerging, or unknown infectious agents cannot be completely excluded.Because the operator sustained a cut while opening the product, contact between the wound and product contents cannot be ruled out.While no infection, medical treatment, or long-term sequelae have been reported and the likelihood of disease transmission is considered remote, infection resulting from exposure to human-derived materials cannot be definitively excluded.Given the potential for serious deterioration in health should transmission occur, this event is conservatively assessed as meeting the definition of a serious injury.
|