The device was returned to zoll medical corporation for evaluation.A review of the clinical log dated 07/24/2026 identified no critical errors.The device powered on, completed the rescue sequence, remained operational for approximately 25 minutes, and delivered eight shocks at 120 j, 150 j, and 200 j.Patient impedance ranged from 81-93 ohms.Given that the device automatically analyzes the patient's ecg and determines whether a shockable rhythm is present, the operator could not control when shocks were advised; therefore, the customer report could not be duplicated.The device performed as expected and designed in this setting.Functional testing with test batteries was completed successfully, and inspection found no device faults.The unit will be recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|