It was reported by the customer that during use, the cardiosave intra-aortic balloon pump (iabp) had a ¿fan fault detected ¿ switch to another maquet iabp, if available, and contact maquet technical support¿ message.Patient was involved, no harm reported.The treatment was completed.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported that the fan located above the motor had an abnormal vibration.As the fans had been replaced in the field action, it was decided that both fans would be replaced with the new ones.After the replacement, the equipment cycled for more than 4 hours and did not present any further messages or any abnormalities in its operation.Product passed all functional and safety tests per manufacturer specifications.
|