• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. ULTRASONIC GASTROVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. ULTRASONIC GASTROVIDEOSCOPE Back to Search Results
Model Number GF-UE160-AL5
Medical Device Problem Codes Dent in Material (2526); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer returned the ultrasonic gastrovideoscope to the service center for a separate issue.During the device analysis, the investigator indicated that a cut on pink rubber and sharp dent on white cap.There was no patient involvement.
 
Additional Manufacturer Narrative
The final investigation has been completed.The failures identified during the inspection has been confirmed.A root cause was traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC GASTROVIDEOSCOPE
Common Device Name
ULTRASONIC GASTROVIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50882329
Report Number3002808148-2026-33610
Device Sequence Number10755020
Product Code ODG
UDI-Device Identifier04953170356261
UDI-Public04953170356261
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K051541
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberGF-UE160-AL5
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/27/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/27/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/12/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-