Medline Industries, LP MEDLINE; IV ADMIN SET, 15DRP, 3 NF PORTS, 112" Back to Search Results
Catalog Number DYNDST1545
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/12/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "our anesthesia provider is having an issue with the blue port leaking on the primary tubing (b)(6).When a secondary line is added for antibiotics, pressure is put on the bag then it pops the blue cap off the vial starts leaking.".
 
Additional Manufacturer Narrative
The account reported finding blue port leaking, and the customer reported that when a secondary line is added for antibiotics, pressure is put on the bag then it pops the blue cap off the vial starts leaking.The customer returned unused samples for review; 15ea samples were randomly selected to be evaluated, an iv administration set was connected to a port on each sample, and fluid flow was initiated through both sets.No leakage was observed at the blue port on any of the samples evaluated, and the complaint is not confirmed.No root cause was identified; the customer reported that a pressure bag was used with the device, and that when a pressure bag wasn't available, manual pressure was applied by squeezing the bag, and per the instructions for use, the device is not pressure-rated.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
IV ADMIN SET, 15DRP, 3 NF PORTS, 112"
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
jarred meyers
three lakes drive
northfield, IL 60093--275
MDR Report Key50882415
Report Number1417592-2026-49685
Device Sequence Number10667565
Product Code FPA
UDI-Device Identifier10884389975961
UDI-Public10884389975961
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K093773
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNDST1545
Device Lot Number0332510A215
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/25/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/12/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-