• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medtronic Mexico S. De R.L. De CV AXIUM PRIME BRPL HLX; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medtronic Mexico S. De R.L. De CV AXIUM PRIME BRPL HLX; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number APB-3-6-HX-ES
Medical Device Problem Codes Separation Failure (2547); Detachment of Device or Device Component (2907); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/21/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received a report of a axium coil that had prematurely detached from non-detachment with the backup method.The patient was undergoing surgery for treatment of a saccular, ruptured aneurysm located in the middle cerebral artery bifurcation section with a max diameter of 5mm and a 2.4mm neck diameter.It was noted the patient's blood flow was normal, and the vessel tortuosity was normal.No images are available for review.It was reported that when using the coil involved in this event, after coil deployment was completed, detachment was attempted using the backup detachment method.After breaking the backup detachment point and pulling out the detachment wire, it was found that the coil had not separated from the delivery wire.The delivery wire was repeatedly advanced and withdrawn, but the coil still could not be detached.Torque was applied, and the delivery wire was repeatedly rotated 360 degrees, but it still could not be detached.The operator then attempted to withdraw the entire coil into the microcatheter.While withdrawing it into the microcatheter, the coil detached.The operator believed that the coil detachment point was abnormal and requested factory inspection.It was indicated that all devices were prepared as per the instructions for use (ifu), and no patient symptoms or further complications were reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM PRIME BRPL HLX
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
Medtronic Mexico S. De R.L. De CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX  22210
Manufacturer (Section G)
Medtronic Mexico S. De R.L. De CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX   22210
Manufacturer Contact
justin ellis
9775 toledo way
irvine, CA 92618
+1(763)526-5677
MDR Report Key50882692
Report Number9617601-2026-05770
Device Sequence Number21279904
Product Code KRD
UDI-Device Identifier00763000313968
UDI-Public(01)00763000313968(17)280901(10)232124185
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K151447
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAPB-3-6-HX-ES
Device Catalogue NumberAPB-3-6-HX-ES
Device Lot Number232124185
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2026
Initial Report FDA Received Date09/24/2026
Date Device Manufactured09/01/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age87 YR
Patient SexFemale
-
-