(b)(4).Date sent: 9/24/2026.D4 batch # 715d85.Investigation summary.The product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the ntlc75 device was received with the staple height selector broken and no returned for analysis, the metal anvil was noted no properly seated and with no reload present.Further analysis showed the anvil was noted partially detached from 3 anvil posts and this area appears to be damaged as if the anvil was pulled out off the device.The event reported was confirmed and due to the condition of the staple height selector and the anvil, the reported complaint was confirmed by an external cause as improper handling.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure that the device meets the required specifications prior to shipment.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 715d85, and no non-conformances were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent:
is the current patient status known?
was there any patient consequence or change in the post-operative care of the patient as a result of the event?.
|