The customer reported receiving a medline glucose evencare g2 that has low readings, with readings almost everyday below 20 or giving out err4 error, and one reading was 48 but the remaining were low.Medline received the evencare g2 blood glucose meter, the sample came with 2 aaa working batteries, and the sample was tested using evencare g2 glucose control solutions and the customer's evencare g2 blood glucose test strips by following the instructions for use.After testing, medline experienced no issues, the sample turned on and off as intended, and control solutions results were within range; the complaint was not confirmed.No root cause was identified, and no trends were identified for the specified material number and its corresponding defect type.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|