| Catalog Number |
0042-0040-US |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The complainant reported that while performing preventative maintenance a crack purge disk retainer was found and the purge film was damaged.The purge disk retainer and purge disk film were replaced.The retainer was torqued to specification and failed measurement at three mmhg spec 0 mmhg +/- 1.After replacing the purge pressure sensor, all specifications were passed.No patient was involved.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by abiomed inc, or its employees that the report constitutes an admission that the product, abiomed inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|