The customer reported that the item was being used when a patient burn occurred, and the side of electrode melted and exposed a hole in the insulation and noticed a small crack leading to the base of the tip; according to the facility, "the patient experienced a burn on the outside of her breast" while using a cautery blade.Medline quality received samples for evaluation, and visual and functional inspection was conducted; visual inspection of the provided defective sample found that the side of the electrode did show signs of melting and a crack was observed at the base of the tip.The reported concern was confirmed, as the sample showed signs of being defective.The suspected root cause is undetermined at this time; the division will continue to monitor this issue for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|