| Model Number |
ENF-VH |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the videoscope to the service center for evaluation of a separate issue.During the evaluation, a reportable malfunction was identified: missing a-rubber.The affected component required replacement.No patient involvement was associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection
based on the results of the investigation, the probable cause of this complaint is expected or random component failure resulting from physical stress, with no design or manufacturing issue identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.B3: olympus detected the issue during the device servicing; therefore, there is no information regarding the customer reported event date.
|
| |
|
Search Alerts/Recalls
|