• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. RHINO-LARYNGO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. RHINO-LARYNGO VIDEOSCOPE Back to Search Results
Model Number ENF-VH
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The customer returned the videoscope to the service center for evaluation of a separate issue.During the evaluation, a reportable malfunction was identified: missing a-rubber.The affected component required replacement.No patient involvement was associated with this event.
 
Additional Manufacturer Narrative
The device was returned to olympus for inspection based on the results of the investigation, the probable cause of this complaint is expected or random component failure resulting from physical stress, with no design or manufacturing issue identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.B3: olympus detected the issue during the device servicing; therefore, there is no information regarding the customer reported event date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RHINO-LARYNGO VIDEOSCOPE
Common Device Name
RHINO-LARYNGO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50883355
Report Number3002808148-2026-33628
Device Sequence Number18830080
Product Code EOB
UDI-Device Identifier04953170310447
UDI-Public04953170310447
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K221638
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberENF-VH
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/27/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/27/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/30/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-